Sys*Stim 206 Instruction Manual—
Rev. E_4/4/13
17
b) Use Mettler Electronics electrodes to ensure safe and
effective operation.
c) Verify connection of the line cord to a grounded wall
receptacle.
d) Connect the electrode cable (ME 2260) into the electrode
jack.
e) Note: Descriptions of the symbols used on controls are in
Section 2.
4.3 Operating
Instructions
1. Set the Intensity control to reset (minimum).
2. Set the Pulse frequency control to Minimum.
3. Insert treatment cords into electrode pads and insert the dual cord
plug into the Sys*Stim 206 electrode jack. To assure conductivity
across the electrode-to-patient interface, saturate electrode sponges
with tap water or other conductive fluids (except when using carbon
flex Tens electrodes where the use of conductive gel is required).
4. Depress the Main power control to the On position. The following
indications should be present (if not, refer to the troubleshooting guide
at the end of this manual.):
Narrow pulse: green indicator lit.
Timer indicator: decimal point only.
Output meter: “0”
Peak volts: green indicator lit.
5. Place electrode pads over area to be treated using proper technique.
(Note: To reduce skin-to-pad resistance, strap the sponge electrode
firmly to the skin without impairing circulation.) Note: No straps are
necessary when using self-adhesive electrodes.
6. Enter the treatment time using the Treatment Time selector.
(Treatment time must be entered using two digits.)
a. For times up to 9 minutes, enter “0” followed by the second digit.
(Example: For 9 minutes, enter “0” then “9”.)
b. For 10 to 29 minutes, enter a “1” or “2” followed by the second
digit. (Example: For 15 minutes, enter a “1” then “5”.)
c. If an incorrect time is entered accidentally, or to change the
entered time, depress the Timer bypass switch (a row of dashed
lines should appear in the indicator) and reselect the desired
treatment time.
For combination ultrasound / stimulation treatment only – depress
Timer bypass to allow a straight 29-minute count up in even seconds.
The Sonicator treatment time selector is used for determining