Sys*Stim 206 Instruction Manual—
Rev. E_4/4/13
13
Section 3—Installation
3.1 Installation
Instructions
1. Plug the attached line cord into a grounded wall outlet that is rated at
120 VAC (±10%)/60Hz
(220 VAC/50Hz for export models)
. Your
power supply must match the voltage requirements listed on the serial
number label of your device.
Do not connect the Sys*Stim 206 to a
power supply rated differently than that described above.
2. The line cord comes equipped with a standard 3–prong plug. This plug
provides grounding for the Sys*Stim 206. Do not defeat its purpose by
using 3–to–2 prong adapters or any other means of attaching to a wall
outlet.
3. Plug the electrode cable (ME 2260) into the electrode jack.
4. The
Sys*Stim 206 may be susceptible to interference originating from
shortwave diathermy units operating in close proximity to it. Avoid
operating the Sys*Stim 206 adjacent to and simultaneously with
operating shortwave devices.
5. Once you have verified proper functioning of your Sys*Stim 206, using
the instructions in Section 4, please register the warranty for your
Sys*Stim 206 on line at
http://www.mettlerelectronics.com/product-
registration/
.