1.2 Defibrillator Analyzer
Specifications
1.
Energy Output Measurement
High Range
Voltage
<5000 volts
Maximum current
120 amperes
Maximum energy
1000 Joules
Accuracy
±
2 % of reading for >100 Joules
±
2 Joule of reading for <100 Joules
Trigger level
100 volts
Playback amplitude
1 mV/1000 V Lead I
Test pulse
100 + 4 Joules
Low Range
Voltage
<1000 volts
Maximum current
24 amperes
Maximum energy
50 Joules
Accuracy
±
2% of reading for >20 Joules
±
2 Joule of reading for <20 Joules
Trigger level
20 volts
Playback amplitude
1 mV/200 V Lead I
Test pulse
Approx. 4 Joules
Load Resistance
50 ohms
±
1%, non-inductive (<1
µ
H)
Display Resolution
0.1 Joules
Measurement Time
Window
100 ms
Absolute Max. Peak
Voltage
6000 volts
Pulse Width
100 ms
Cardioversion
Measured time delay
±
2 ms
Oscilloscope Output
High measure range
1000:1 amplitude-attenuated
Low measure range
200:1 amplitude-attenuated
Waveform Storage And Playback
Discharge can be viewed via ECG outputs and paddles.
Output: 200:1 Time Base expansion.
Sync Time Measurements
Timing window
Starts - 40 ms at each R-wave peak.
Test waveforms
All waveform simulations available.
Delay time accuracy
±
1 ms
Charge Time Measurement
From 0.1 seconds to 99.9 seconds.
2
Содержание QA-45
Страница 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Страница 5: ...This page intentionally left blank v...
Страница 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Страница 26: ...This page intentionally left blank 9...
Страница 28: ...2...
Страница 38: ...4...
Страница 42: ...This page intentionally left blank 8...
Страница 58: ...This page intentionally left blank 10...
Страница 60: ...A 2...
Страница 61: ...A 3...
Страница 62: ...A 4...
Страница 63: ...A 5...
Страница 64: ...A 6...
Страница 65: ...A 7...
Страница 66: ...A 8...
Страница 67: ...A 9...
Страница 68: ...A 10...
Страница 69: ...A 11...
Страница 70: ...A 12...
Страница 71: ...A 13...