Select the desired rhythm by pressing
UP (F2) or DOWN (F3)
.
Save this under ‘Wave” in the STATUS field by pressing
Se-
lect
. Press
CANCEL (F5)
to cancel selection. The ECG signal
is output through the low-level ECG connectors, high-level
ECG connector, and paddle contact plates on the QA-45.
4. Set the defibrillator to analyze the ECG rhythm and operate in
the automatic and semi-automatic mode.
5. Records the defibrillator’s response.