TP8 - TP6
+8.8V
±
0.2V
TP8 - TP4
-8.0V
±
0.3V
TP8 - TP2
+2.5V
±
0.075V
TP8 - TP5
-8V
+2 -1V
TP8 - TP3
+5V
±
0.1V
5. Connect the frequency counter to TP7 and read the frequency.
Requirement: 2 MHz (
±
0.002 MHz).
6. Slowly reduce the voltage from the power supply until the in-
stant the display gives the message: 'Change battery, and reset
system'. Measure the operating voltage with the multimeter.
Requirement: 6.9V (
±
0.3v). Return the operating voltage to
9.0V. and reset QA-45 by switching off the power for a short
period.
7. Select 80 BPM in the ECG Wave Menu. Connect the oscillo-
scope to TP1 for signal, and TP8 to ground. Check that QA-45
generates a 80 BPM signal with an amplitude of approximately
250 mV on the R pulse. Connect the oscilloscope to the High
Level ECG contact and check that the same signal is present,
only with an amplitude of approximately 1V on the R pulse.
8. Set power to OFF. Measure the resistance from the RL output to
the RA, LA and LL outputs.
Requirement: 1000 ohm (
±
30 ohm).
9. Measure the resistance from the RL output to the V2, V3, V4,
V5 and V6 outputs.
Requirement: 1000 ohm (
±
30 ohm).
10. Set power to ON and go into the
SYSTEM TEST (F1)
of the
page 3
menu and choose ECG + by moving the cursor and
pressing
SELECT (F4)
. Measure and adjust the voltage be-
tween TP1 and TP8 to 0.490V
±
1 mV. Choose ECG 0 from
menu. Measure the voltage between TP1 and TP8 and check
that it is b20 mV and -10 mV.
11. Choose ECG + from the System Test Menu, measure the volt-
ages between the ECG outputs, and check that they fall within
the limits shown in the following table:
Contact
Power limit
Nom. Value
Upper limit
RL - RA
1.20 mV
1.35 mV
1.50 mV
RL - LA
2.40 mV
2.65 mV
2.90 mV
RL - LL
3.00 mV
3.35 mV
3.70 mV
RL - V1
2.70 mV
3.02 mV
3.30 mV
RL - V2
3.30 mV
3.72 mV
4.10 mV
RL -V3
4.00 mV
4.49 mV
5.00 mV
RL - V4
4.50 mV
5.06 mV
5.60 mV
RL- V5
4.00 mV
4.49 mV
5.00 mV
RL - V6
3.30 mV
3.72 mV
4.10 mV
12. Measure the voltage at the High ECG output.
Requirement: 2.0V (
±
0.05v).
2
Содержание QA-45
Страница 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Страница 5: ...This page intentionally left blank v...
Страница 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Страница 26: ...This page intentionally left blank 9...
Страница 28: ...2...
Страница 38: ...4...
Страница 42: ...This page intentionally left blank 8...
Страница 58: ...This page intentionally left blank 10...
Страница 60: ...A 2...
Страница 61: ...A 3...
Страница 62: ...A 4...
Страница 63: ...A 5...
Страница 64: ...A 6...
Страница 65: ...A 7...
Страница 66: ...A 8...
Страница 67: ...A 9...
Страница 68: ...A 10...
Страница 69: ...A 11...
Страница 70: ...A 12...
Страница 71: ...A 13...