1. Introduction
This chapter describes the Metron’s QA-45 Defibrillator /
Transcutaneous Pacemaker Analyzer features and specifications.
1.1
QA-45 Features
The QA-45 Analyzer is a precision instrument for testing defibrilla-
tors and transcutaneous pacemakers, and is designed to be used by
trained service technicians.
The defibrillator function of the QA-45 measures the energy output,
and ensures that the defibrillator complies with specified require-
ments. QA-45 has a built-in load resistance of 50 ohm, which
roughly corresponds to the impedance of the human body. The de-
fibrillator pads are placed on the QA-45 contact plates. Thus, the de-
fibrillator is connected through the load resistance. When the defib-
rillator is discharged, QA-45 calculates and displays the energy de-
livered.
In the pacer function the QA-45 tests all types of transthoracic pace-
makers. The testing is menu driven, and simple to operate. QA-45
measures and displays a pulse’s amplitude, rate, energy and width. It
also conducts demand sensitivity tests, measuring and displaying re-
fractory periods, and immunity tests, which determine the pacemak-
er’s susceptibility to 50/60 Hz interference.
1
Содержание QA-45
Страница 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Страница 5: ...This page intentionally left blank v...
Страница 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Страница 26: ...This page intentionally left blank 9...
Страница 28: ...2...
Страница 38: ...4...
Страница 42: ...This page intentionally left blank 8...
Страница 58: ...This page intentionally left blank 10...
Страница 60: ...A 2...
Страница 61: ...A 3...
Страница 62: ...A 4...
Страница 63: ...A 5...
Страница 64: ...A 6...
Страница 65: ...A 7...
Страница 66: ...A 8...
Страница 67: ...A 9...
Страница 68: ...A 10...
Страница 69: ...A 11...
Страница 70: ...A 12...
Страница 71: ...A 13...