4.
ADV. ALG.
(Advisory Algorithms)
(F2)
.
These ECG algorithms are meant to test the analysis and prompt-
ing feature of automatic and semi-automatic defibrillators.
Choose desired selection by pressing
UP (F2)
or
DOWN (F3)
.
Save this under ‘Wave” in the STATUS field by pressing
SELECT (F4)
. Press
CANCEL (F5)
to cancel selection.
5.
CHARGE TIME (F3).
Used to test the battery and charging
capacitor in the defibrillator. It changes the text ‘Delay’ to
‘Chrg T’ in the RESULT field in the main menu.
6.
PRINT HEADER (F4).
Automatically writes a heading for the
new test protocol.
7.
WAVE AMPL. (Wave Amplitude) (F1).
Choose desired amplitude by pressing
UP (F2)
or
DOWN (F3)
.
Save this under ‘Ampl” in the STATUS field by pressing
SELECT (F4)
. Press
CANCEL (F5)
to cancel selection.
8.
PLAY PULSE (F2)
enables playback of the last discharge.
9.
PERF. WAVE (Performance ECG) (F3).
Choose desired wave by pressing
UP (F2)
or
DOWN (F3)
. Save
this under ‘Wave” in the STATUS field by pressing
SELECT
(F4)
. Press
CANCEL (F5)
to cancel selection.
10.
SYSTEM TEST (F1) .
5
Содержание QA-45
Страница 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Страница 5: ...This page intentionally left blank v...
Страница 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Страница 26: ...This page intentionally left blank 9...
Страница 28: ...2...
Страница 38: ...4...
Страница 42: ...This page intentionally left blank 8...
Страница 58: ...This page intentionally left blank 10...
Страница 60: ...A 2...
Страница 61: ...A 3...
Страница 62: ...A 4...
Страница 63: ...A 5...
Страница 64: ...A 6...
Страница 65: ...A 7...
Страница 66: ...A 8...
Страница 67: ...A 9...
Страница 68: ...A 10...
Страница 69: ...A 11...
Страница 70: ...A 12...
Страница 71: ...A 13...