5. Transcutaneous Pacemaker Mode Testing

This chapter explains QA-45 transcutaneous, or pacer mode testing,

5.1

Introduction

QA-45 tests all types of transthoracic pacemakers. The testing is 

menu driven, and simple to operate. QA-45 measures and displays a 
pacer pulse’s amplitude, rate, energy and width. It also conducts de-

mand sensitivity tests, measuring and displaying refractory periods, 
and immunity tests, which determine the pacemaker’s susceptibility 

to 50/60 Hz interference.

5.2 Testing Preparation 

1. Connect the pacer output cables to the pacer input connectors.

2. Switch the mode switch to ‘PACE’ mode.

3. Turn the QA-45 on. The following will be displayed in the LCD 

display for about two seconds:

4. The following main menu will then appear:

5. Press 

SELECT LOAD (F1)

.  The following load options will 

appear:

The load range is 50 to 2300 ohms in steps of 50 ohms up to 200 

ohms, and 100 ohms from 200 up to 2300 ohms

1

Содержание QA-45

Страница 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...

Страница 2: ...anty METRON warrants that the QA 45 Defibrillator Transcutaneous Pacemaker Analyzer will substan tially conform to published specifications and to the documentation provided that it is used for the pu...

Страница 3: ...Printouts 8 4 Defibrillator Mode Testing 1 4 1 Introduction 1 4 2 Test Preparation 1 4 3 Energy Test 3 4 4 Cardioversion Test 4 4 5 Maximum Energy Charging Time Test 6 4 6 Shock Advisory Algorithm Te...

Страница 4: ...ative authorized to process changes and revisions to publications make the pen changes or insert the revised pages ensure that obsolete pages are withdrawn and either disposed of immediately or marked...

Страница 5: ...This page intentionally left blank v...

Страница 6: ...resistance of 50 ohm which roughly corresponds to the impedance of the human body The de fibrillator pads are placed on the QA 45 contact plates Thus the de fibrillator is connected through the load...

Страница 7: ...ack amplitude 1 mV 200 V Lead I Test pulse Approx 4 Joules Load Resistance 50 ohms 1 non inductive 1 H Display Resolution 0 1 Joules Measurement Time Window 100 ms Absolute Max Peak Voltage 6000 volts...

Страница 8: ...Playback 200 to 1 time base expansion of defibrillator pulse by playback to ECG Leads Manual ECG Performance Test DC Pulse 4 seconds 1 0 mV Square wave 2 Hz 1 0 mV p p biphasic Triangular wave 2 Hz 1...

Страница 9: ..._STD PVC PVCRonT R on T PVC mfPVC Multifocal PVC bigeminy Bigeminy run5PVC Bigeminy Run of 5 PVCs vtach Ventricular Tachycardia Shock Advisory Test Algorithms ASYS Asystole SVTa_90 Supraventricular Ta...

Страница 10: ...Accuracy 1 or 2 ppm Pulse width 0 6 to 80 ms Accuracy 1 or 0 3 ms 3 DEMAND SENSITIVITY TEST Waveforms Square SQR Triangle TRI and Havemine SSQ ECG output Amplitude 0 4 mV Resolution 40 V Pacer input...

Страница 11: ...8 0 cm 11 0 in Weight 2 06 kg with battery 4 5 lbs Recommended Printer HP DeskJet 500C 550C and Canon BJ 10SX Standard Accessories 110 V or 220 V AC Adapter P N 17021 Internal paddle contact adapter P...

Страница 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...

Страница 13: ...ch a replacement The customer will not have to wait for a claim to be investigated by the supplier The customer should place a new purchase order to ensure delivery 4 When returning an instrument to M...

Страница 14: ...remain off for at least 5 seconds before switching on again in order to allow the test cir cuits to discharge fully 2 Low Battery Power If battery power falls below 6 9 volts 0 3 volts the display wil...

Страница 15: ...e instrument 3 Changing Batteries Open the compartments in the base of the instrument replace the old batteries with new ones and close the compartment covers Use 9 volt alkaline batteries Duracell MN...

Страница 16: ...zer It allows you to conduct the same tests but by remote control via an IBM compatible PC XT with MS Windows Version 3 1 or later Additionally the program has additional fea tures to enhance your def...

Страница 17: ...stored on disk and the results of testing can be sent back to the equipment database to close a work order and update the service his tory 5...

Страница 18: ...een PACE and Low High ranges of defibrilla tor energy 3 LCD Display Shows messages test results and function menus 4 Function Keys Fl F5 are used to select the functions shown on the bottom line of th...

Страница 19: ...oded 4 mm safety terminals with snap to banana adapters 7 Pacer Input Connectors The pacer output cables are connected to these so that the pacer signal passes through the instrument with a variable l...

Страница 20: ...t for connecting 9V 100 mA battery elimina tor 3 2 QA 45 Menu and Function Keys The QA 45 uses display and programmable function keys to provide flexibility and control over the operations The upper p...

Страница 21: ...the QA 45 has been switched on 2 Main Menu a Main Menu Bar Page 1 Mode switch in Low or High position b Second Menu Bar Page 2 c Third Menu Bar Page 3 3 ECG WAVE F1 Choose desired wave by pressing UP...

Страница 22: ...ator It changes the text Delay to Chrg T in the RESULT field in the main menu 6 PRINT HEADER F4 Automatically writes a heading for the new test protocol 7 WAVE AMPL Wave Amplitude F1 Choose desired am...

Страница 23: ...soft ware PRO Soft QA 40 45 3 4 Menu and Messages Transcutaneous Pacemaker Mode 1 Startup Screen The following screen will be displayed for 2 seconds after the QA 45 has been switched on 2 Main Menu a...

Страница 24: ...ress CANCEL F5 to cancel selection 4 SELECT NOISE F2 Choose desired noise for the immunity test by UP F2 or DOWN F3 and then SELECT F4 Press CANCEL F5 to cancel selection 5 PRINT HEADER F3 Automatical...

Страница 25: ...ensitive to any external inputs the maximum time interval after the generation of a pacer pulse and maximum time interval after a QRS wave See Pacing Refractory Period and Sensing Refractory Period in...

Страница 26: ...This page intentionally left blank 9...

Страница 27: ...gy is equal to the square of the voltage divided by the load resistance QA 45 measures and records the voltage pulse every 100 s 1000 times for a total time of 100 ms The squares of the voltages are t...

Страница 28: ...2...

Страница 29: ...QA 45 calculates the energy delivered over the load resistance and displays the result in joules under RESULT See below QA 45 also shows the energy measured the maximum voltage and the maximum current...

Страница 30: ...ect ECG WAVE F1 from the main menu 2 The ECG Wave menu opens QA 45 includes the following ECG wave selection for cardioversion tests or for the testing of elec trocardiograph monitors Normal Sine Rate...

Страница 31: ...under Wave in the STATUS field by pressing SE LECT F4 Press CANCEL F5 to cancel selection 3 QA 45 includes the following ECG wave amplitude options 0 5 mV 1 0 mV 1 5 mV and 2 0 mV To change wave ampl...

Страница 32: ...ew pulse can be generated when the LCD display shows LOCAL 4 5 Maximum Energy Charging Time Test 1 The charge time function is used to test the battery and the charging capacitor in the defibrillator...

Страница 33: ...matic and semi au tomatic defibrillators A series of arrhythmia is available for analysis by the defibrillator that should then prompt the user to shock of no shock in accordance with national and int...

Страница 34: ...lect Press CANCEL F5 to cancel selection The ECG signal is output through the low level ECG connectors high level ECG connector and paddle contact plates on the QA 45 4 Set the defibrillator to analyz...

Страница 35: ...displaying refractory periods and immunity tests which determine the pacemaker s susceptibility to 50 60 Hz interference 5 2 Testing Preparation 1 Connect the pacer output cables to the pacer input co...

Страница 36: ...Select the desired noise form by pressing UP F2 or DOWN F3 and then Select F4 Press CANCEL F5 to cancel the selection After selection the main menu will reappear 2...

Страница 37: ...amplitude mV re quired to cause the pacemaker to operate in the demand mode During sensitivity measurement three different waveforms are selectable with widths varying in steps from 10 to 200 ms This...

Страница 38: ...4...

Страница 39: ...ctory Period Test 1 General This test is used to test the time interval in millisec onds ms during which the pacemaker is insensitive to any ex ternal inputs QA 45 does this by measuring the maximum t...

Страница 40: ...Period The maximum time interval after a QRS wave is sensed by the pacemaker during which time the presence of a second QRS wave is ignored The sensed refractory period is measured in a similar man ne...

Страница 41: ...c Upon completion of testing the results will be displayed un der RESULT Press REF PER CANCEL F5 to cancel the test 7...

Страница 42: ...This page intentionally left blank 8...

Страница 43: ...erences The function keys are numbered 1 to 5 with switch 1 farthest to left 6 4 Test 1 Set power to ON Wait 3 seconds and measure the current from the power supply Requirement 68 mA 5 mA 2 Adjust P2...

Страница 44: ...FF Measure the resistance from the RL output to the RA LA and LL outputs Requirement 1000 ohm 30 ohm 9 Measure the resistance from the RL output to the V2 V3 V4 V5 and V6 outputs Requirement 1000 ohm...

Страница 45: ...values will be stored in QA 45 s EEPROM when SAVE QUIT is pressed 15 Set the power switch to Off Remove the covers at J106 and J107 Measure the resistance between TP101 and TP102 Requirement 2 MOhm 2...

Страница 46: ...value displayed is between 1997 5 and 2000 mV regardless of the polarity status Check that the scope output is 2000 mV 20 mV or 2000 mV 20 mV Remove the power source and secure P102 P103 and P104 Swi...

Страница 47: ...0 ohm 297 ohm 303 ohm 400 ohm 396 ohm 404 ohm 500 ohm 495 ohm 505 ohm 600 ohm 594 ohm 606 ohm 700 ohm 693 ohm 707 ohm 800 ohm 792 ohm 808 ohm 900 ohm 892 ohm 909 ohm 1000 ohm 990 ohm 1010 ohm 1100 ohm...

Страница 48: ...iation Rate 80 ppm 0 5 ppm Width 10 ms 0 3 ms Amplitude 20 mA 1 mA Energy 2 0 mJ 0 5 mJ 31 Set the pacer load to 1000 ohm Connect a pulse generator to the pacer input The following values are acceptab...

Страница 49: ...a printer A serial port RS 232C makes remote control from a PC possible All measurement and control signals are con nected via contacts on the top and rear of the unit To operate the unit there are 5...

Страница 50: ...e the external power supply can be connected to J7 D1 protects against incorrect polarization R16 is a PTC fuse that provides high impedance if too much power is drawn F19 provides high frequency filt...

Страница 51: ...4 IC112 reads parallel data from the switches used for operation The data format is adapted to the microprocessor s SPI interface and clocked by the SCK clock from the processor The A D converter IC10...

Страница 52: ...Range With an input attenuation of 200 the measurement range becomes 1000 to 1000 Volt Low Range With an input attenuation of 10 the measurement range becomes 50 to 50V pacer mode IC102 is used as an...

Страница 53: ...ching type and will therefore use no power from the batteries except when they are operated Before the relays are set they are all reset via pin 18 on U4 2 Stimulus Output The stimulus signal that is...

Страница 54: ...de BAT54 2 D7 D8 Zener diode 13V 1 D1 Transzorbdiode BZW06 6V4B 1 D3 Transzorbdiode BZW06 17B 3 D9 Dll Crystal 8 MHz 1 Y1 Resistor 1K0 1 0 5W 1 R8 Resistor 2K2 1 0 5W 1 R9 Resistor 3K3 1 0 5W 1 R13 Re...

Страница 55: ...4052P 1 ICl14 Diode 1N4148 3 Dlll Dl13 Zener diode 5V1 4 D104 D107 Schottky diode 1N5819 1 Dl14 Transzorb diode BZW06 10B 2 D101 D102 Transzorb diode BZW06 17B 3 D108 Dl10 Resistor 47R 1 0 5W 1 R184 R...

Страница 56: ...5 nylon 4 Screw M3x12 nylon 4 Nut M3 nylon 4 ECG SIGNAL DISTR BOARD ECG signal distr circuit board AR 094 1 Resistor 499R 1 0 5W 10 R2 R3 R5 R7 R9 Rll R13 R15 R17 R19 Resistor 47K5 1 0 5W 1 R14 Resist...

Страница 57: ...r 10K 1 0 5W 3 R15 R24 R27 Resistor 20K 0 1 0 5W 4 R8 R10 R14 R16 Resistor 24K 0 1 0 5W 2 R17 R18 Resistor 30K 0 1 0 5W 2 R26 R29 Resistor 249K 0 1 0 5W 2 R19 R20 Multilay X7R 100nF 50V 8 C1 C5 C7 C8...

Страница 58: ...This page intentionally left blank 10...

Страница 59: ...Board A 5 Sensor Board Component Location Diagram A 6 Schematic Diagram Part 4 Sensor Board A 7 Schematic Diagram Part 5 Sensor Board A 8 Schematic Diagram Part 6 Sensor Board A 9 ECG Signal Distribu...

Страница 60: ...A 2...

Страница 61: ...A 3...

Страница 62: ...A 4...

Страница 63: ...A 5...

Страница 64: ...A 6...

Страница 65: ...A 7...

Страница 66: ...A 8...

Страница 67: ...A 9...

Страница 68: ...A 10...

Страница 69: ...A 11...

Страница 70: ...A 12...

Страница 71: ...A 13...

Страница 72: ...s Fax E mail Date Error Report Product Version Serial no Description of the situation prior to the error Description of the error METRON AS internally Comments Received date Correction date Ref No Cri...

Страница 73: ...ame Phone Address Fax E mail Date Improvement Suggestion Product Version Description of the suggested improvement METRON AS internally Comments Received date Correction date Ref No Critical Normal Min...

Страница 74: ...6 Fax 1 616 454 3350 E mail support us metron biomed com FRANCE 30 rue Paul Claudel 91000 Evry France Phone 33 1 6078 8899 Fax 33 1 6078 6839 E mail info metron fr NORWAY Travbaneveien 1 N 7044 Trondh...

Отзывы: