4.3 Energy Test
1. Select a suitable energy range using the mode switch.
•
Use the HIGH range for normal adult testing.
•
Use the LOW range for low energy testing, where the en-
ergy does not exceed 50 Joule and the peak voltage does
not exceed 1200 volts.
2. Securely place the defibrillator paddles on the QA-45 contact
plates, and discharge the defibrillator. The APEX (+) pad
should be connected to the right-hand plate, and the STERNUM
pad to the left plate. This ensures correct signal polarity for the
oscilloscope output. A reversal of this configuration will not
damage the QA-45, nor will it give incorrect energy readings.
However, the polarity of the oscilloscope output will simply be
reversed. The discharge from the defibrillator is transferred to
the QA-45's load resistance.
3. QA-45 calculates the energy delivered over the load resistance
and displays the result in joules under RESULT. See below:
QA-45 also shows the energy measured, the maximum voltage
and the maximum current in the energy wave. Following the
discharge from the defibrillator, QA-45 shows a playback of the
wave from the ECG output. A new pulse can be generated when
the LCD display shows 'LOCAL'.
4. Following a discharge from the defibrillator, the instrument
shows a playback of the wave from the ECG output. The dis-
play will thus be in playback mode. When this is shown in one
line, QA-45 automatically prints out the result.
5. The discharged pulse can be repeated. To do this press
more-2
(F5)
to advance to page 2 of the main menu.
Press
PLAY PULSE (F2)
. The display will show 'Oper: Play-
back,' and displays the result in joules under RESULT.
APEX (+) pad
→
right plate
STERNUM pad
→
left plate
Note
If the maximum voltage for a se-
lected range is exceeded, the LCD
display will show ‘WARNING!
Overload’
3
Содержание QA-45
Страница 1: ...QA 45 User Service Manual QA 45 Defibrillator and Transutaneous Pacemaker Analyzer P N 13060...
Страница 5: ...This page intentionally left blank v...
Страница 12: ...Periodic Inspection The unit should be calibrated every 12 months 7...
Страница 26: ...This page intentionally left blank 9...
Страница 28: ...2...
Страница 38: ...4...
Страница 42: ...This page intentionally left blank 8...
Страница 58: ...This page intentionally left blank 10...
Страница 60: ...A 2...
Страница 61: ...A 3...
Страница 62: ...A 4...
Страница 63: ...A 5...
Страница 64: ...A 6...
Страница 65: ...A 7...
Страница 66: ...A 8...
Страница 67: ...A 9...
Страница 68: ...A 10...
Страница 69: ...A 11...
Страница 70: ...A 12...
Страница 71: ...A 13...