f.
Release pressure and verify that the display reads 0, if
not: power off and repeat from step 2
.
g.
Check the calibration and linearity by applying pressure of
330 mmHg (440 mBar) and 165 mmHg (220 mBar), and
verify that the QA-1290 displays this value
±
1mmHg.
Pressure sensor
Reference
Measured value
0 mBar (0 mmHg)
220 mBar (165 mmHg)
440 mBar (330 mmHg)
660 mBar (495 mmHg)
*Tolerance Pressure Sensor:
±
1 mmHg
5. The calibration procedure is now complete.
7-2