The connection of a NIBP Analyzer and cuff that use two tubes is as
shown below. Tubing from the NIBP Analyzer is connected to the
luer lock on the Inlet Port, while the Cuff’s tubing is connected to the
luer lock on the Cuff Port.
With its Tubing Kit (P.N. 17515) METRON offers a number of pres-
sure connector adapters. The adapters connect to the QA-1290 Pres-
sure Port via a quick-disconnect, airtight, O-ring pressure fitting.
See
Appendix A, Tubing Kit Components.
Note:
The QA-1290 has an internal fixed volume that is used to simu-
late the cuff once the simulation begins . By connecting the inlet port
and cuff port of the QA-1290 together and then to the monitor under
test testing can be done without an external cuff connected.