
>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>
>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>
>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>
>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>>
At the end of the transfer, a normal startup sequence will occur.
4.
Test to confirm the upgrade.
•
Exit the DOS program and unhook the RS-232 transfer ca-
ble from the QA-1290.
•
Switch the QA-1290
OFF
, then turn it back
ON
.
Verify that the new firmware version is installed while the QA-
1290 is running its start sequence. The firmware version will be
displayed in Startup Screen Menu 3.
3-8
Содержание QA-1290
Страница 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Страница 6: ...This page intentionally left blank vi...
Страница 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Страница 30: ...This page intentionally left blank 4 8...
Страница 36: ...This page intentionally left blank 5 6...
Страница 51: ...Appendix A Tubing Kit Components A 1...
Страница 52: ...This page intentionally left blank B 2...
Страница 54: ...B 2...
Страница 55: ...B 3...
Страница 56: ...B 4...
Страница 57: ...B 5...
Страница 58: ...B 6...
Страница 59: ...B 7...
Страница 60: ...B 8...
Страница 61: ...B 9...
Страница 62: ...This page intentionally left blank B 10...