MAIN MENU (F4)
to return to Main Menu 1.
Manometer: 0 mmHg ZERO PRESSURE.> F7
Peak: 0 mmHg RESET PEAK ...> F6
PRINT ........> F5
INFLATE SET UP MAIN MENU
F1 F2 F3 F4
5. Press
INFLATE (F2)
. The menu changes and
DEFLATE (F2)
becomes active
(see below).
Wait until the inflation stops. When
the test results are shown, press
DEFLATE
.
Manometer: 0 mmHg ZERO PRESSURE.> F7
Peak: 0 mmHg RESET PEAK ...> F6
PRINT ........> F5
DEFLATE SET UP MAIN MENU
F1 F2 F3 F4
6. To reset the manometer and peak values to 0 mmHg, press
ZERO
PRESSURE (F7)
and
RESET PEAK (F6)
. Note that this also
resets the time to zero.
7. If you want to print the test results, press
PRINT (F5)
to advance
to the Print Screen.
5.3 Manometer
This test will let you select a pressure in mmHg and maintain it.
1. Advance to Main Menu Screen 2. From this screen press
Ma-
nometer (F6)
. This advances you to the test screen.
Equipment Code : ......................> F7
Leak Test .............................> F6
BP Test ...............................> F5
MORE SET UP
F1 F2 F3 F4
Over-Pressure Cut-Off Test ............> F7
Manometer .............................> F6
Print header/footer ...................> F5
GO BACK SET UP
F1 F2 F3 F4
5-4
Содержание QA-1290
Страница 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Страница 6: ...This page intentionally left blank vi...
Страница 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Страница 30: ...This page intentionally left blank 4 8...
Страница 36: ...This page intentionally left blank 5 6...
Страница 51: ...Appendix A Tubing Kit Components A 1...
Страница 52: ...This page intentionally left blank B 2...
Страница 54: ...B 2...
Страница 55: ...B 3...
Страница 56: ...B 4...
Страница 57: ...B 5...
Страница 58: ...B 6...
Страница 59: ...B 7...
Страница 60: ...B 8...
Страница 61: ...B 9...
Страница 62: ...This page intentionally left blank B 10...