5.2 Over-Pressure Cut-Off
Test
1. Advance to Main Menu Screen 2. From this screen press
Over-
Pressure Cut-Off Test (F7)
. This advances you to the test
screen.
Equipment Code : ......................> F7
Leak Test .............................> F6
BP Test ...............................> F5
MORE SET UP
F1 F2 F3 F4
Over-Pressure Cut-Off Test ............> F7
Manometer .............................> F6
Print header/footer ...................> F5
GO BACK SET UP
F1 F2 F3 F4
2. Press
SET UP (F3)
. This advances you to the Inflate Pressure
Screen.
(See below).
Manometer: 0 mmHg ZERO PRESSURE.> F7
Peak: 0 mmHg RESET PEAK ...> F6
PRINT ........> F5
INFLATE SET UP MAIN MENU
F1 F2 F3 F4
Inflate Pressure:XXX mmHg F7
F6
F5
STORE GO BACK MAIN MENU
F1 F2 F3 F4
3. The
Inflate Pressure (F7)
value must be entered. Otherwise the
test will fail. Use the alphanumeric keys to enter the pressure set-
ting to between 0 and 500 mmHg. Press the
Enter
key to save
the new setting(s). Use the
Delete
key to clear the current setting.
4. Press:
•
STORE (F1)
to store the settings in flash memory.
•
GO BACK (F2)
to return to the Over-Pressure Cut-Off Test
Screen.
5-3
Содержание QA-1290
Страница 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Страница 6: ...This page intentionally left blank vi...
Страница 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Страница 30: ...This page intentionally left blank 4 8...
Страница 36: ...This page intentionally left blank 5 6...
Страница 51: ...Appendix A Tubing Kit Components A 1...
Страница 52: ...This page intentionally left blank B 2...
Страница 54: ...B 2...
Страница 55: ...B 3...
Страница 56: ...B 4...
Страница 57: ...B 5...
Страница 58: ...B 6...
Страница 59: ...B 7...
Страница 60: ...B 8...
Страница 61: ...B 9...
Страница 62: ...This page intentionally left blank B 10...