1-8
Gain/Damping:
2 Hz square wave
Frequency
Response
Low Frequency:
4 second DC pulse
Band Pass:
10 Hz sine
Monitor:
-3dB point: 40 Hz sine
Power Line Notch Filter:
50 Hz sine
Linearity:
2 Hz triangle wave
5.
ECG Lead Test
Display flashes if lead resistance is <3 kOhms (DC lead wire only)
6.
Blood Pressure General
Input/Output Impedance:
300 Ohms
Exciter Voltage Range:
2 to 16 volts
Exciter Frequency Range:
DC to 4000 Hz
Output Sensitivity:
5 or 40 µV/V/mmHg
Accuracy:
± 1% full scale
1 mmHg
Calibrated Rate:
80 BPM
7.
Dynamic Blood Pressure Selections
Atmosphere:
0 mmHg
Arterial:
120/80 mmHg
Left Ventricle:
20/0 mmHg
Right Ventricle:
25/0 mmHg
Pulmonary Artery:
25/10 mmHg
Pulmonary Wedge:
10/2 mmHg
Static Levels:
0, 20, 40, 80, 100, 200, 250, 300 mmHg
8.
Arrhythmia
Selections
Atrial Fibrillation 1 & 2
Premature Atrial Contraction
Premature Ventricular Contraction (PVC)
Early
PVC
R on T PVC
Multifocal
PVCs
Bigeminy
Bigeminy Run of 5 PVCs
Ventricular
Tachycardia
Ventricular
Fibrillation
Second Degree Type 2
Right Bundle Branch Block
Asynchronous
Pacemaker
Fetal/Maternal
ECG
Содержание PS-416M
Страница 1: ...1 1 PS 416M User Service Manual PATIENT SIMULATOR...
Страница 4: ...1 4 This page intentionally left blank...
Страница 6: ...1 6 This page intentionally left blank...
Страница 10: ...1 10 This page intentionally left blank...
Страница 12: ...2 12 This page intentionally left blank...
Страница 20: ...4 20 This page intentionally left blank...
Страница 27: ...5 27 APPENDIX A DIAGRAMS Component Location Diagram Schematic Diagram Part 1 Schematic Diagram Part 2...
Страница 28: ...5 28 This page intentionally left blank...
Страница 29: ...5 29 Component Location...
Страница 30: ...5 30 Schematic Diagram Part 1...
Страница 31: ...5 31 Schematic Diagram Part 2...
Страница 32: ...5 32 This page intentionally left blank...
Страница 34: ...5 34 This page intentionally left blank...
Страница 36: ...5 36...