4 Installation
IC equipment: Inline Ultrafiltration 2 - push mode
■■■■■■■■
17
■
Tighten the PTFE capillary (6.1803.040) to the outlet
OUT
of the
sample chamber using a PVDF pressure screw (6.2744.000).
■
Tighten the PTFE capillary (6.1803.050) to the outlet
OUT
of the
filtrate chamber using a PVDF pressure screw (6.2744.000).
6 Connecting the remaining capillaries
■
Tighten the free end of the PEEK capillary (6.1831.160) to the
sample needle using a pressure screw (6.2744.014) (
see the man-
ual for the sample changer
).
■
Tighten the free end of the PTFE capillary (6.1803.040) to Port 1
of the injector using a pressure screw (6.2744.014) (
see the man-
ual for the ion chromatograph
).
■
Either tighten the free end of the PTFE capillary (6.1803.040) to
the waste collector or guide it directly to the waste container and
tighten it there using a PVDF pressure screw (6.2744.000).
4.3
Inserting the ultrafiltration cell
Fastening the ultrafiltration cell to the sample changer
Accessories
■
Filtration cell holder (6.02057.030)
■
Ultrafiltration cell 2 (6.02729.110)
1 Attaching the holder
Tighten the filtration cell holder (6.02057.030) to the sample changer
(
see manual for the sample changer
).
2 Inserting the ultrafiltration cell