4.2 Connecting the ultrafiltration cell
■■■■■■■■■■■■■■■■■■■■■■
14
■■■■■■■■
IC equipment: Inline Ultrafiltration 2 - push mode
4.2
Connecting the ultrafiltration cell
This chapter describes how to establish the capillary connections in the fil-
tration system. This chapter does not describe how the peristaltic pumps
are connected. Information on the installation of the peristaltic pumps can
be found in the
chapter "Installing the peristaltic pump" in the manual
for the ion chromatograph or in the manual for the sample changer
.
The following figure shows a schematic overview of all capillary connec-
tions in the filtration system.
Figure 5
push mode – Connecting the ultrafiltration cell
NOTICE
■
In order to keep dead volume to a minimum, the capillaries of the IC
equipment: Inline Ultrafiltration 2 - push mode are as short as possi-
ble.
■
To prevent the capillaries leading into the ion chromatograph from
being pinched, always guide them through the capillary feed-
throughs provided for this purpose (
see the manual for the ion chro-
matograph
).
Connecting the ultrafiltration cell
Accessories
■
Ultrafiltration cell 2 (6.02729.110)
■
PTFE capillary, 0.5 mm ID / 20 cm (6.1803.050)
■
2 x PTFE capillary 0.5 mm ID / 1 m (6.1803.040)
■
Pump tubing LFL (orange/yellow), 3 stoppers (6.1826.320)
■
Pump tubing LFL (yellow/yellow), 3 stoppers (6.1826.390)