![Metrohm 6.02729.110 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/metrohm/6-02729-110/6-02729-110_manual_1779546021.webp)
■■■■■■■■■■■■■■■■■■■■■■
4 Installation
IC equipment: Inline Ultrafiltration 2 - push mode
■■■■■■■■
13
Using the tweezers, place the wet filtration membrane within the cir-
cular engraving on the sample chamber.
7 Assembling the ultrafiltration cell
Place the filtrate chamber on the sample chamber in such a way that
the 2 guide bolts of the sample chamber fit into the bore holes of the
filtrate chamber.
8 Screwing the ultrafiltration cell together
■
Screw the screws entirely into the ultrafiltration cell using a hex
key.
■
Additionally, tighten each screw by a
1
/
4
rotation.
9
If the ultrafiltration cell is not used immediately, seal the sample inlet,
the sample outlet and the filtrate outlet with the 3 green stoppers.
This allows for the filtration membrane to stay moist.