4.2 Connecting the ultrafiltration cell
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
IC equipment: Inline Ultrafiltration 2 - push mode
4 Connecting the capillaries to the pump tubing for the filtrate
6.1803.050
6.2744.070
6.2744.070
6.1803.040
6.2744.034
6.2744.160
6.1826.320
■
Tighten the PTFE capillary (6.1803.050) to the inlet of the pump
tubing with orange/yellow stoppers (6.1826.320) using a pressure
screw (6.2744.070).
■
Tighten the PTFE capillary (6.1803.040) to the outlet of the pump
tubing with orange/yellow stoppers (6.1826.320) using a pressure
screw (6.2744.070).
5 Connecting capillaries to the ultrafiltration cell
CAUTION
Damage to the ultrafiltration cell caused by using the
wrong pressure screws
PEEK pressure screws are too hard. PEEK pressure screws can dam-
age the ultrafiltration cell material.
■
Use only transparent PVDF pressure screws (6.2744.000) to
tighten capillaries to the ultrafiltration cell.
6.2744.000
6.2744.000
6.2744.000
6.1803.050
6.1803.050
6.1803.040
■
Tighten the PTFE capillary (6.1803.050) to the inlet
IN
of the sam-
ple chamber using a PVDF pressure screw (6.2744.000).