3 Installation
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■■■■■■■■
51
Replacing the sample loop
The sample loop can be replaced, depending on requirements. For addi-
tional information concerning selection of the appropriate sample loop,
.
NOTE
Use only 6.2744.010 PEEK pressure screws for connecting capillaries
and sample loop to the injection valve.
1 Removing existing sample loop
■
Loosen 6.2744.010 pressure screws at connector 3 and connec-
tor 6.
■
Remove sample loop.
2 Mounting new sample loop
■
Fasten one end of the sample loop
with a 6.2744.010
PEEK pressure screw
to connector 3.
■
Fasten the other end of the sample loop
with a second
6.2744.010 PEEK pressure screw
to connector 6.
3.16.2
Mode of operation of the injection valve
The injection valve
tions -
FILL
and
INJECT
. Switching back and forth between the two valve
positions determines whether the sample path or the eluent path is gui-
ded through the sample loop. The following figure provides a schematic
display of the flow paths of the two valve positions.
1
2
4
5
3
6
1
2
4
5
3
6
1
2
4
3
5
4
3
1
2
A
B
Figure 28
Injection valve – Positions
A
Position FILL
B
Position INJECT