3.3 Installation of the low pressure gradient
■■■■■■■■■■■■■■■■■■■■■■
16
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
3.3
Installation of the low pressure gradient
The installation of the 850 Professional IC – Anion – MSM-HC – MCS – LP
Gradient must be adjusted slightly in order to be operated as a low pres-
sure gradient system. The following installation instructions list the neces-
sary work steps in brief. At each step you will find cross-references to
more detailed installation instructions for individual components, should
you require such aids.
Installing the low pressure gradient
1 Expanding the instrument to low pressure gradient
■
Loosen the 6.1834.090 connection tubing from the output of the
eluent degasser.
■
Dismount the coupling
and pump head inlet capillary
of the high pressure pump.
■
Use the short pressure screw to tightly screw the 6.2758.020 mix-
ing coil for low pressure gradients directly onto the input of the
high pressure pump,
and the lon-
ger pressure screw to screw it tightly to the output of the mixing
valve
(see "Connecting mixing coil for low pressure gradients",
■
Screw on the 6.1834.120 connection tubing to the output of the
eluent degasser
(see "Connecting the eluent degasser", page
and connect with input A of the mixing valve
ing connection tubings", page 41)
■
Equip two 6.1834.080 eluent aspiration tubings
eluent aspiration tubing", page 33)
ent degasser inputs in the low pressure gradient
the eluent degasser", page 37)
.