3 Installation
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
■■■■■■■■
49
1
2
3
4
Figure 26
Sample degasser
1
Sample degasser input
2
Sample degasser output
3
PEEK pressure screw, long 6.2744.090
4
Connection capillaries 6.1803.040
Connecting the sample degasser
1
Remove and keep 6.2744.220 thread stoppers from the input and
output of the sample degasser.
2
Connect the end of the 6.1803.040 sample aspirating capillary con-
nected to the injection valve with a long PEEK pressure screw
to the output of the sample degasser
3
Connect 6.1803.040 connection capillary with a long PEEK pressure
screw
to the input of the sample degasser
.
4
Guide the other end of the connection capillary out of the instrument
through a capillary feed-through.
CAUTION
If the sample degasser is not used, the input and output
must
be
sealed with the 6.2744.220 thread stoppers.