3.7 Rear of the instrument
■■■■■■■■■■■■■■■■■■■■■■
26
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
NOTE
Up to two detectors can be positioned and connected.
CAUTION
The instrument
must
be turned off when connecting a detector.
1 Removing rear panel
■
Unscrew knurled screws
on the rear panel.
■
Remove handle if still fastened to the instrument.
■
Remove rear panel
2 Positioning the detector
■
Put the detector through this opening and position it on the sup-
port surface intended for this purpose. Slide it right up to the
front.
3 Replacing rear panel
■
Insert the detector cable in a cable feed-through
on the rear
panel
.
■
Place the detector output capillary in a suitable capillary feed-
through.
■
Replace
(Optionally, the handle can be remounted higher up and used as
a holder for MPaks.)
■
knurled screws.
4 Connecting the detector
NOTE
The instrument has two detector connection sockets
Detec-
tor 1
and
Detector 2
. You must ensure that the selected connec-
tor corresponds to the connector entered in the MagIC Net
method.
Recommendation
: Use
Detector 1
as standard. In the AnCat sys-
tem with 2 detectors:
Detector 1
for anions,
Detector 2
for cati-
ons.