5.6 Inline filter
■■■■■■■■■■■■■■■■■■■■■■
92
■■■■■■■■
850 Professional IC–Anion – MSM-HC – MCS – LP Gradient
Changing the filter
The flow must be stopped before changing the filter.
1 Removing the inline filter
■
Unscrew the pressure screws
from the inline filter.
2 Unscrewing the filter screw
■
Screw the filter screw
the aid of two 6.2621.000 adjustable wrenches.
3 Inserting the filter
■
with tweezers.
■
flat in the filter housing with tweezers
.
4 Mounting filter screw
■
Screw the filter screw
and tighten by hand. Then additionally tighten slightly with two
6.2621.000 adjustable wrenches.
5 Remounting the inline filter
■
Screw the pressure screws
back onto the inline filter.
6 Rinsing the inline filter
■
Dismantle the guard column (if present) and the separation col-
umn and replace with a 6.2744.040 coupling.
■
Rinse the instrument with eluent.