background image

Remote programming 

 

100 000 cts digital multimeter 

7   

Multimeter specific 

commands

 

(cont’d)

 

 

 

[SENSe:]  

 

 

CLAMP:CAMP1Ratio[?] 

reads or sets the numerator of the ratio 
used in measuring amps on the clamp 
function.  

 

CLAMP:CAMP2Ratio[?] 

reads or sets the denominator of the 
coefficient used in measuring amps on 
the clamp function.  

 

CLAMP:CUNit[?] 

reads or sets the unit displayed in clamp 
measurement. 

 

CLAMP:CVOLT2Ratio[?] 

reads or sets the numerator of the ratio 
used to measure volt in clamp function.  

 

CLAMP:CVOLT1Ratio[?] 

reads or sets the denominator of the 
ratio used to measure volt-ampere in 
clamp function.  

 

CLAMP:MEASure[?] 

reads or sets the measurement used in 
clamp function. 

 

CLAMP:Status[?] 

reads or sets the activity status of the 
clamp mode.  

 

FILTer[:LPASs][:Status][?] 

reads or sets the activity status of the 
meter filter.  

 

FREQuency:MODe[?] 

reads or sets the measurement range of 
the frequency. 

 FREQuency:THReshold:VO

LTage:RANGe[?] 

reads or sets the voltage range for 
frequency measurement.  

 

FUNCtion[?] 

reads or sets the type of main measurem. 

 

HOLD:Status[?] 

reads or sets the status of HOLD mode. 

 

MENU:DBM:IMPedance[?] 

reads or sets the dBm impedance 
calculation. 

 MENU:WATT:IMPedance[?] 

reads 

or 

sets the calculation impedance 

of power measurement.  

 

RANGe:AUTO[?] 

reads or sets the activity status of the 
multimeter autorange mode. 

 

RANGe:AUTO:PEAK[?] 

reads or sets the status of the multimeter 
autoranging mode on peak values .  

 

RANGe[:UPPer] [?] 

reads or selects the range of the main 
measurement.  

 

SECondary[?] 

reads or sets the group of second. meas.

 

TEMPerature:TRANsducer[?] 

reads or sets the type of sensor used for 
measuring temperature. 

SYSTem  

 

 :BEEPer:Status[?] 

 

reads/sets the activation status of the 
beeper.  

 :COMMunicate:SERial[:REC

eive]:BAUD[?] 

reads or sets the communication speed 
of the serial link.  

 

:DATE[?] 

reads or sets the date. 

 

:ERRor[:NEXT]? 

reads the error number.  

 

:LANGuage[?] 

reads or sets the language of the menus. 

 

:LOCal 

sets the meter in local mode. 

 

:PROTocole 

selects the MODBUS protocol.  

 

:TIME[?] 

reads or sets the time.  

 

:VERSion? 

reads the SCPI release. 

TRACe  

 

 

:CATalog? 

see DATA command: CATalog  

 

[:DATA]:VALue? 

see DATA command: [DATA:] VALue?  

 

:DELete:ALL 

see DATA command: DELete: ALL  

 

:DELete[NAME] 

see DATA command: DELete [NAME]  

 

:POINts[?] 

see DATA command: POInts [?]  

 

:RATE[?] 

see command DATA: RATE [?]  

 

:STOre:Status[?] 

see DATA command: STOre: Status [?]  

UNIT  

 

 

:TEMPerature[?] 

reads or sets the unit of meas. of temp. 

 

 
 

Содержание ASYC IV MTX 3292

Страница 1: ...Portable colour graphic multimeters ASYC IV 100 000 cts MTX 3292 MTX 3293 Remote programming Pôle Test et Mesure CHAUVIN ARNOUX ...

Страница 2: ...e PC Configuration Communication 9600 19200 34800 baud 8 data bits 1 stopbit no parity no flow control Remote control The programming instructions comply with standard IEEE488 2 protocol SCPI They provide the user with the possibility of checking the instrument remotely from simple standard controls Communication between a controller and a generator enables users to Configure the instrument Perfor...

Страница 3: ...ombination of both The defined type parameters are marked by the characters The brackets mean that the parameters are optional The vertical bar may be read as an or it is used to separate the various possible parameters Separators The separator descends in the next directory or returns under the root if preceded by a The separator separates two commands in the same directory or marks the end of a ...

Страница 4: ...sitioned at root level When the analyzer is in a directory both separators are necessary to return to the root IEEE 488 2 common commands Commands Functions CLS erases the event status registers and the error list ESE reads or sets the event authorization ESR reads the event status register IDN reads the identifier OPC reads or sets the command synchronizing RST re initializes the instrument SRE r...

Страница 5: ...ts the function of the secondary display FUNCtion LIST reads the list of authorized functions of the secondary display MATH MAFactor reads or sets the A coefficient of the mathematical function of the running measurement MATH MBFactor reads or sets the B coefficient of the mathematical function of the running measurement MATH MUNit reads or sets the unit of the mathematical function of the running...

Страница 6: ... of the device RATE reads or sets the logging interval STOre Status reads or sets the status for the recording mode DISPlay LUMInosity reads or sets the brightness of the display HELP reads the list of commands HCOPy DEVice CMAP reads the color of the display SDUMp IMMediate reads the contents of the bitmap of the display INPut COUPling reads or sets the type of running coupling volt or ampere mea...

Страница 7: ...HOLD mode MENU DBM IMPedance reads or sets the dBm impedance calculation MENU WATT IMPedance reads or sets the calculation impedance of power measurement RANGe AUTO reads or sets the activity status of the multimeter autorange mode RANGe AUTO PEAK reads or sets the status of the multimeter autoranging mode on peak values RANGe UPPer reads or selects the range of the main measurement SECondary read...

Страница 8: ...e value decimal number between 0 and 255 which represents the event validation registry value ESR Standard Event Status Register Query To the ESR question the instrument returns the event status registry value Query syntax ESR Response format value T value Decimal number between 0 and 255 which represents the event validation registry value IDN Identification Number Query To the IDN question the i...

Страница 9: ... syntax SRE Response format value T The SRE command sets the service request activation registry Command syntax SRE value value Decimal number between 0 and 255 which represents the value of the service request activation register STB Read Status byte Query To the STB question the instrument returns the value of the status byte Query syntax STB Response format value T value Decimal number between ...

Страница 10: ...memorized in surv mode Response format 2014 08 24 3 23 49 AVERage DATE STAR Query To the question CALC AVER DATE START the instrument returns the date and time when the surveillance mode has started Response format 2014 08 24 3 23 49 AVERage DATE STOP Query To the question CALC AVER DATE STOP the instrument returns the date and time when the surveillance mode has stopped Response format 2014 08 24...

Страница 11: ...H MAF factor sets the B coefficient of the mathematical function of the running measurement To the question CALC MATH MAF the instrument returns the B coefficient of the mathematical function of the running measurement The factor parameter is a decimal numeric value between 9 9999e 99 and 9 9999e 99 MATH MUNit Command Query The command CALC MATH MUN unit sets the unit of the mathematical function ...

Страница 12: ... question CALC SPEC PERC the instrument returns the specification value in percentage of the running measurement SPEC SMAX Query To the question CALC SPEC SMAX the instrument returns the value of the upper limit of the specification for the running measurement SPEC SMIN Query To the question CALC SPEC SMIN the instrument returns the value of the lower limit of the specification for the running mea...

Страница 13: ...e measurement campaign Names of available measurement campaigns can be known using the DATA CAT command POINts Command Query The command DATA POIN points sets the buffer size of the memorization unit To the question DATA POIN the instrument returns the buffer size of the memorization unit The parameter points is an integer value RATE Command Query The command DATA RATE cadence sets the recording c...

Страница 14: ...r is the size of the next field which itself contains the number of binary bytes that are transmitted SDUMp IMMediate Query To the question HCOP SDUM the instrument returns the content of the display bitmap The first character is The second character is the size of the next field which itself contains the number of binary bytes that are transmitted Commands INPut COUPling Command Query The command...

Страница 15: ...rator is a numeric decimal value from 0 1e 6 to 9999 0e 6 CLAMP CUNit Command Query The command CLAMP CUN unit sets the unit of the clamp function To the question CLAMP CUN the instrument returns the unit of the clamp function The parameter unit is a chain of 3 characters maximum If the character chain is OHM the unit displayed is the symbol Ω CLAMP CVOLT1Ratio Command Query The command CLAMP CVOL...

Страница 16: ... activate To the question FILT the instrument returns the filter activation status FREQuency MODe Command Query The command FREQ MOD INF200KHZ SUP200KHZ selects the frequency measurement range To the question FREQ MOD the instrument returns used frequency measurement range FREQuency THReshold VOLTage RANGe Command Query The command FREQ THR VOLT RANG range selects the voltage range of the frequenc...

Страница 17: ...ce for the power measurement The parameter impedance is an integer value from 1 to 10000 RANGe AUTO Command Query The command RANG AUTO 0 1 OFF ON activates or deactivates the autorange of the active main measurement The values are 0 or OFF to de activate and 1 or ON to activate To the question RANG AUTO the instrument returns the activation status of the autorange RANGe AUTO PEAK Command Query Th...

Страница 18: ...od mathematical function 2 Frequency dB mathematical function 3 dBm dBm reference mathematical function 4 Peak positive negative peak crest factor 5 Power reference power mathematical function 6 VA current mathematical function 7 Period positive duty cycle mathematical function 8 Period negative duty cycle mathematical function 9 Positive pulse time positive pulse counter mathematical function 10 ...

Страница 19: ...ERial RECeive BAUD Command Query The command SYST COMM SER BAUD 9600 19200 38400 selects the communication speed of the serial link device To the question SYST COMM SER BAUD the instrument returns the communication speed of the serial link device DATE Command Query The command SYST DATE year month day sets the date To the question SYST DATE the instrument returns the date The parameter year is an ...

Страница 20: ... to be set to 1 101 Invalid character 103 Invalid separator 104 Data type error 108 Parameter not allowed 109 Missing parameter 111 Header separator error 112 Program mnemonic too long 113 Undefined header 114 Header suffix out of range 121 Invalid character in number 128 Numeric data not allowed 141 Invalid character data 148 Character data not allowed 151 Invalid string data 154 String data too ...

Страница 21: ... hour minute second sets the time To the question SYST TIME the instrument returns the time The parameter time is an integer value from 0 to 23 The parameter minute is an integer value from 0 to 59 The parameter second is an integer value from 0 to 59 VERSion Query To the question SYST VERS the instrument returns the SCPI version it supports Response includes the year and the revision index Respon...

Страница 22: ...in Arnoux GmbH Slamastrasse 29 2 4 1230 Wien Tel 01 61 61 961 0 Fax 01 61 61 961 61 SCANDINAVIA CA Mätsystem AB Sjöflygvägen 35 SE 18304 TÄBY Tel 46 8 50 52 68 00 Fax 46 8 50 52 68 10 SCHWEIZ Chauvin Arnoux AG Einsiedlerstraße 535 8810 Horgen Tel 044 727 75 55 Fax 044 727 75 56 UNITED KINGdOM Chauvin Arnoux Ltd Waldeck House Waldeck Road Maidenhead SL6 8BR Tel 01628 788 888 Fax 01628 628 099 MIDDL...

Отзывы: