Remote programming
100 000 cts digital multimeter
15
Command READ?
(Query)
To the question
READ?
, the instrument returns the result of the running
measurement.
Response format
: +276.91
mVAC
Commands [SENSe:]
CLAMP:CAMP1Ratio[?]
(Command/Query)
The command
CLAMP:CAMP1 <numerator>
selects the numerator of
the ratio used in amp measurement in the clamp function.
To the question
CLAMP:CAMP1?
, the instrument returns the ratio
numerator used in amp measurement in the clamp function.
The parameter <numerator> is a numeric decimal value from +0.1
e-6
to
+9999.0
e+6
.
CLAMP:CAMP2Ratio[?]
(Command/Query)
The command
CLAMP:CAMP2 <denominator>
selects the
denominator of the ratio used in amp measurement in the clamp
function.
To the question
CLAMP:CAMP2?
, the instrument returns the
denominator of the ratio used in amp measurement in the clamp
function.
The parameter <numerator> is a numeric decimal value from +0.1
e-6
to
+9999.0
e+6
.
CLAMP:CUNit[?]
(Command/Query)
The command
CLAMP:CUN <unit>
sets the unit of the clamp function.
To the question
CLAMP:CUN?
, the instrument returns the unit of the
clamp function.
The parameter <unit> is a chain of 3 characters maximum.
If the character chain is "OHM", the unit displayed is the symbol "
Ω
".
CLAMP:CVOLT1Ratio[?]
(Command/Query)
The command
CLAMP:CVOLT1 <numerator>
selects the numerator
of the ratio used in volt measurement in clamp function.
To the question
CLAMP:CVOLT1?
, the instrument returns the
numerator of the ratio used in volt measurement in the clamp function.
The parameter <numerator> is a numeric decimal value from +0 to
+9999.0
e+6
.
CLAMP:CVOLT2Ratio[?]
(Command/Query)
The command
CLAMP:CVOLT2 <denominator>
selects the
denominator of the ratio used in volt measurement in clamp function.
To the question
CLAMP:CVOLT2?
, the instrument returns the
denominator of the ratio used in volt measurement in clamp function.
The parameter <numerator> is a numeric decimal value from +0 to
+9999.0
e+6
.