Remote programming
12
100 000 cts digital multimeter
REFerence:Status[?]
(Command/Query)
The command
CALC:REF:STAT <0|1|OFF|ON>,
activates or de-activates
the relative mode.
The values are 0 or OFF to de-activate and 1 or ON to de-activate.
To the question
CALC:REF:STAT?,
the instrument returns the activation
status of the relative mode.
SPEC:DIGITs?
(Query)
To the question
CALC:SPEC:DIGIT?
, the instrument returns the
specification value in digit of the running measurement.
SPEC:PERCent?
(Query)
To the question
CALC:SPEC:PERC?
, the instrument returns the
specification value in percentage of the running measurement.
SPEC:SMAX?
(Query)
To the question
CALC:SPEC:SMAX?
, the instrument returns the value of
the upper limit of the specification for the running measurement.
SPEC:SMIN?
(Query)
To the question
CALC:SPEC:SMIN?
, the instrument returns the value of
the lower limit of the specification for the running measurement.
SPEC:Status[?]
(Command/Query)
The command
CALC:SPEC:STAT <0|1|OFF|ON>,
activates or de-
activates the specification mode.
The values are 0 or OFF to de-activate and 1 or ON to activate.
To the question
CALC:SPEC:STAT?,
the instrument returns the activation
status of the specification mode.