Remote programming
100 000 cts digital multimeter
17
HOLD:Status[?]
(Command/Query)
The command
HOLD:STAT <OFF|ON|AUTO>,
sets the HOLD mode
of the instrument.
To the question
HOLD:STAT?,
the instrument returns the HOLD mode
status.
MENU:DBM:IMPedance[?]
(Command/Query)
The command
MENU:DBM:IMP <impedance>,
sets the calculation
impedance of the dBm measurement.
To the question
MENU:DBM:IMP?
, the instrument returns the value of
the selected calculation impedance for the dBm measurement.
The parametere <impedance> is an integer value from 1 to 10000.
MENU:WATT:IMPedance[?]
(Command/Query)
The command
MENU:WATT:IMP <impedance>,
sets the calculation
impedance of the power measurement.
To the question
MENU:WATT:IMP?
, the instrument returns the value
of the selected calculation impedance for the power measurement.
The parameter <impedance> is an integer value from 1 to 10000.
RANGe:AUTO[?]
(Command/Query)
The command
RANG:AUTO <0|1|OFF|ON>
activates or deactivates
the autorange of the active main measurement.
The values are 0 or OFF to de-activate and 1 or ON to activate.
To the question
RANG:AUTO?
, the instrument returns the activation
status of the autorange.
RANGe:AUTO:PEAK[?]
(Command/Query)
The command
RANG:AUTO:PEAK <0|1|OFF|ON>
activates or de-
activates the autorange on peak values of the active main
measurement.
The values are 0 or OFF to de-activate and 1 or ON to activate.
To the question
RANG:AUTO?
, the instrument returns the activation
status of autoranging on peak values.
RANGe[:UPPER][?]
(Command/Query)
The command
RANG <range>
selects the range of the active main
measurement.
To the question
RANG ?,
the instrument returns the number of the
range of the active main measurement.
The parameter <range> is a numeric decimal value that defines the
range.