Remote programming
16
100 000 cts digital multimeter
CLAMP:MEASure[?]
(Command/Query)
The command
CLAMP:MEAS <measure>
selects the measure used
by the clamp function.
To the question
CLAMP:MEAS?
, the instrument returns the mesure
used by the clamp function.
The parameter <measure> is a chain of characters which is either
"VOLTAGE", or "CURRENT".
CLAMP:Status[?]
(Command/Query)
The command
CLAMP:STAT <0|1|OFF|ON>,
activates or de-activates
the clamp function.
The values are 0 or OFF to de-activate and 1 or ON to activate.
To the question
CLAMP:STAT?,
the instrument returns the activation
status of the clamp function.
FILTer[:LPASs][:Status][?]
(Command/Query)
The command
FILT <0|1|OFF|ON>,
activates or inhibits the instrument
filter.
The values are 0 or OFF to de-activate and 1 or ON to activate.
To the question
FILT?,
the instrument returns the filter activation status.
FREQuency:MODe[?]
(Command/Query)
The command
FREQ:MOD <INF200KHZ|SUP200KHZ>,
selects the
frequency measurement range.
To the question
FREQ:MOD?,
the instrument returns used frequency
measurement range.
FREQuency:THReshold:
VOLTage:RANGe[?]
(Command/Query)
The command
FREQ:THR:VOLT:RANG <range>,
selects the voltage
range of the frequency measurement.
To the question
FREQ:THR:VOLT:RANG?,
the instrument returns the
voltage range used for the frequency measurement.
The parameter <range> is a decimal numeric value which defines the
range.
If the parameter <range> has a negative value,
the range change in
frequency measurement is automatic.
FUNCtion[?]
(Command/Query)
The command
FUNC <"VOLTage" | " CURRent " | " RESistance " | "
FREQuency " | " CONTinuity " | " DIODe " | "50 ohm" | " CAPAcitor
" | " TEMPerature " | "LOWZvoltage" | "DIODEZ" | "PREF">
selects
the main measurement type.
To the question
FUNC?,
the instrument returns the active main
measurement type.