Multi LAN 350
Operation
64
Memory screen – location
with already stored tests.
An occupated memory
location.
It is possible to overwrite
the data that are already
stored.
The user will be asked for
confirmation (YES).
Saving procedure - general
After performing an Complete Autotest / Near End Autotest / Remote Autotest / Fiber
optic Autotest press SAVE key.
The MEMORY menu is displayed.
The following actions can now be performed:
Left, Right:
change the CABLE, OBJECT or FLOOR number (the actual item is
underlined, number is flashing). If a result under the selected
location already exists a warning is displayed.
Up, Down:
switch between CABLE, OBJECT or FLOOR
SAVE:
save the Autotest result in the selected location. If the location is
occupated the user will be asked for confirmation.
ESC:
cancel the saving procedure
After the saving procedure the previous menu is displayed.
Note:
The Autotest results are stored in a FLASH memory that doesn't need any battery
backup. The data will not be lost even if the batteries are removed from the instrument.
5.11.3 Downloading data / results to PC
Refer to chapters 5.13.13 and 1.
5.12 Remote unit MultiLAN 350 RU operation
The Remote unit is used to correctly terminate the far cable end during different
measurements. It executes commands received from the Main unit (instrument). Seven
LEDs indicate the current unit status.
The meaning of keys and LEDs are described in chapter 2.2.2.
Testing mode
While executing measurement commands from the Main Unit the
TESTING
LEDs are
lighting.
Switching the unit On Off