Multi LAN 350
Operation
57
ESC:
returns to the Test Standard screen. New settings are not
considered
ENTER:
the set configuration is stored. The instrument returns to Test
Standard screen
Up, Down:
select measurement
Left, Right:
set measurement On (Yes) or Off (No).
Autotest sequence screen
Note:
All three screens must be confirmed with ENTER to accept the new setup!
The user can recall the default settings in the Settings menu or by setting the Autotest
sequence according to data in Appendix 2 (the tests with limits applied are set On, other
Off).
5.8 Setting optical fiber cable standard
Refer to chapter 5.13.10 for how to select fiber cables measuring mode.
Test standard, fiber cable type an other fiber link parameters can be set in the Test
Standard Menu.
Pass/Fail limits are calculated on basis of selected standard and entered cable data.
The limits and parameters are then used in the Autotest function.
In Appendix 3 a table with preprogrammed test standards and cable types with
associated test and parameter settings can be found.
5.8.1 Optical fiber test standard selection
After Standard Menu is selected with the rotary switch the Test Standard main screen is
displayed. In the upper rows the actual set test standard, cable type, no. of adapters,
no. of splices and length are displayed. The following actions can be performed now:
Up, Down:
select a new test standard
Enter:
enters the Cable Type menu