MI 3152(H)
Eurotest
XC (2,5 kV)
Instrument operation
34
Table 4.5: Maximum test times related to selected test current for general (non-delayed) RCD
Standard
½
I
N
I
N
2
I
N
5
I
N
EN 61008 / EN 61009
500 ms
500 ms 200 ms 150 ms
IEC 60364-4-41
1000 ms 1000 ms 150 ms 40 ms
BS 7671
2000 ms 500 ms 200 ms 150 ms
AS/NZS 3017 (IV)
1000 ms 1000 ms 200 ms 150 ms
Table 4.6: Maximum test times related to selected test current for selective (time-delayed) RCD
1)
Minimum test period for current of ½
I
N
, RCD shall not trip-out.
2)
Test current and measurement accuracy correspond to AS/NZS 3017 requirements.
3)
U
0
is nominal U
LPE
voltage.
Note:
Trip-out limit times for PRCD, PRCD-K and PRCD-S are equal to General (non-delayed)
RCDs.
4.6.5 Initial Settings
In this menu the instrument settings, measurement parameters and limits can be set to initial
(factory) values.
Figure 4.8: Initial settings menu
Warning:
Following customized settings will be lost when setting the instruments to initial settings:
Measurement limits and parameters,
Parameters and settings in General settings menu.
If the batteries are removed the custom made settings will be lost.
Note:
Following customized settings will stay:
Profile settings,
Data in memory.