MI 3109 EurotestPV Lite
Technical specifications
66
9.2 Continuity
9.2.1 Resistance R LOW
Measuring range according to EN61557 is 0.16
1999
.
Measuring range R (
)
Resolution (
)
Accuracy
0.00
19.99
0.01
(3 % of r 3 digits)
20.0
199.9
0.1
(5 % of reading)
200
1999
1
Open-circuit voltage ........................... 6.5 VDC
9 VDC
Measuring current .............................. min. 200 mA into load resistance of 2
Test lead compensation ..................... up to 5
The number of possible tests ............ > 2000, with a fully charged battery
Automatic polarity reversal of the test voltage.
9.2.2 Resistance CONTINUITY
Measuring range (
)
Resolution (
)
Accuracy
0.0
19.9
0.1
(5 % of r 3 digits)
20
1999
1
Open-circuit voltage ........................... 6.5 VDC
9 VDC
Short-circuit current ........................... max. 8.5 mA
Test lead compensation ..................... up to 5