MI 3109 EurotestPV Lite
Technical specifications
70
9.5 General data
Power supply voltage ........................ 9 V
DC
(6
1.5 V battery or accu, size AA)
Operation .......................................... typical 20 h
Charger socket input voltage ............ 12 V
10 %
Charger socket input current ............ 400 mA max.
Battery charging current ................... 250 mA (internally regulated)
Measuring category .......................... 1000 V DC CAT II
600 V CAT III
300 V CAT IV
Protection classification .................... double insulation
Pollution degree ................................ 2
Protection degree ............................. IP 40
Display ............................................ 128x64 dots matrix display with backlight
Dimensions (w
h
d) ..................... 23 cm
10.3 cm
11.5 cm
Weight ............................................ 1.3 kg, without battery cells
Reference conditions
Reference temperature range ........... 10
C
30
C
Reference humidity range ................. 40 %RH
70 %RH
Operation conditions
Working temperature range .............. 0
C
40
C
Maximum relative humidity ............... 95 %RH (0
C
40
C), non-condensing
Storage conditions
Temperature range ........................... -10
C
+70
C
Maximum relative humidity ............... 90 %RH (-10
C
+40
C)
80 %RH (40
C
60
C)
Communication transfer speed:
RS 232 ............................................ 57600 baud
RS 232 wireless ................................ 9600 baud
USB
............................................ 256000 baud
Size of memory:
I-V curve: ......................................... ca. 500 measurements
Other measurements: ...................... ca 1800 measurements
The error in operating conditions could be at most the error for reference conditions
(specified in the manual for each function) +1 % of measured value + 1 digit, unless
otherwise specified in the manual for particular function.