MI 3109 EurotestPV Lite
Measurements
46
If the data is measured with other measuring equipment they can be entered manually.
Select
ENV.
function and
MANUAL
sub-function using the function selector keys and
Up/Down keys.
Keys:
TEST
Enters menu for manual setting of environmental parameters.
Enters menu for changing selected parameter.
Confirms set value of parameter.
/
Selects environmental parameter.
Selects value of parameter.
Function
selector
Exits environmental menu and select PV measurement.
ESC
Exits to main menu.
Exits menu for manual setting of environmental parameters.
Exits menu for changing selected parameter without changes.
Figure 5.33: Example of manually entered results
Displayed results (Irr, Tamb or Tcell are the same as if measured.
Note
:
Environmental parameters are cleared when the instrument is switched Off.
The environmental parameters menu is accessible in Single test and Autotest
operating modes.
5.5.1 Operation with A 1378 PV Remote Unit
See PV Remote Unit User Manual.
5.6 Uoc / Isc measurement
The Uoc / Isc test is intended to check if protection devices in the d.c. part of the PV
installation are effective. The measured data can be calculated to nominal data (STC
values).
See chapter 4.2 Function selection for
instructions on key functionality.
Figure 5.34: Uoc / Isc test