MI 3109 EurotestPV Lite
Measurements
36
Connect
PV continuity test lead to the instrument.
Compensate
test leads resistance (if necessary, see
section 5.2.3
Disconnect
from mains supply and discharge the object to be tested.
Connect
test leads to the tested object (see
Press the
TEST
key to begin performing a continuous measurement.
Press the
TEST
key to stop measurement.
Store
the result by pressing the
MEM
key (optional).
Figure 5.9: Example of continuous resistance measurement
Displayed result:
R............resistance
U:............ actual voltage on test inputs
5.2.3 Compensation of test leads resistance
This chapter describes how to compensate the test leads resistance in both continuity
functions, R LOW
Ω and CONTINUITY. Compensation is required to eliminate the
influence of test leads resistance and the internal resistances of the instrument on the
measured resistance. The lead compensation is therefore a very important feature to
obtain correct result.
symbol is displayed if the compensation was carried out successfully.
Circuits for compensating the resistance of test leads
Figure 5.10: Shorted test leads
Compensation of test leads resistance procedure
Select
R LOW
Ω
or
CONTINUITY
function.
Connect
Continuity PV test lead to the instrument and short the test leads
together (see Figure 5.10).
Press
TEST
to perform
resistance measurement.
Press the
CAL
key to compensate leads resistance.