CS 2890 Power Simulator / Calibrator
Operating the instrument
51
Phase type
– using left and right cursor key, user can switch between Single(L1)
and Poly-phase event type.
Frequency
– using left and right cursor key, user can set signalling frequency in
0.1 Hz increments in range from 50.0 Hz to 3000.0 Hz.
New settings will apply when SET is pressed or when Signalling settings submenu is
closed.
Figure 57: Signalling settings submenu
Table 34: Keys in signalling settings submenu
F4
SET
Refresh (activate) currently set signalling.
Scrolls cursor between options
Modifies parameter.
ENTER
Enters parameter value selection window.
Returns to the Main menu.
Exits from parameter value selection window.
4.10.6
Transient
Transient is overvoltage with a duration of a few milliseconds. Power Simulator
generates oscillatory damped transient on U1 channel, as shown on figure below.
Transient event has overshoot approximately 70% of nominal voltage high and last
about 8% of period duration (period is defined with Frequency parameter), as shown on
figure below.
Содержание CS 2890
Страница 1: ...Power Simulator Calibrator CS 2890 Instruction manual Version 1 4 2 Code No 20 752 950...
Страница 88: ...CS 2890 Power Simulator Calibrator Annex 88...
Страница 89: ...CS 2890 Power Simulator Calibrator Annex 89...
Страница 90: ...CS 2890 Power Simulator Calibrator Annex 90...
Страница 91: ...CS 2890 Power Simulator Calibrator Annex 91...
Страница 92: ...CS 2890 Power Simulator Calibrator Annex 92...
Страница 94: ...CS 2890 Power Simulator Calibrator Annex 94...
Страница 95: ...CS 2890 Power Simulator Calibrator Annex 95...
Страница 96: ...CS 2890 Power Simulator Calibrator Annex 96...