CS 2890 Power Simulator / Calibrator
Operating the instrument
66
6.1.5 Calibration of current inputs for Class S PQA (MI 2883 example)
Reference
value from the CS 2890
UUC
indicated value on the calibration item (MI 2883)
Uncertainty
expanded measurement uncertainty (k=2)
Error
UUC
– Reference
Limit
limit of error (
±), for MI 2883 ±0.5%
UUC set-up: Voltage ratio 1:1, Connection: 4W, Synchronisation: I1, System frequency:
50 Hz, Primary current: 5A, Secondary current: 5A
Reference Frequency
UUC
Uncertainty
Error
Limit
Range=100A
Input
mV
Hz
A
A
A
A
I1
50
50
49.96
0.04
0.04
0.25
I2
50
50
50.02
0.04
-0.02
0.25
I3
50
50
50.04
0.04
-0.04
0.25
N
50
50
50.12
0.04
-0.12
0.25
I1
100
50
99.94
0.07
0.06
0.50
I2
100
50
99.93
0.07
0.07
0.50
I3
100
50
100.0
0.07
0
0.50
N
100
50
100.1
0.07
-0.1
0.50
I1
200
50
199.9
0.15
0.1
1.0
I2
200
50
199.8
0.15
0.2
1.0
I3
200
50
200.0
0.15
0
1.0
N
200
50
200.1
0.15
-0.1
1.0
Range=1000A
I1
50
50
49.97
0.04
0.03
0.25
I2
50
50
50.01
0.04
-0.01
0.25
I3
50
50
50.03
0.04
-0.03
0.25
N
50
50
50.13
0.04
-0.13
0.25
I1
1000
50
1000.0
0.7
0
5.0
I2
1000
50
999.9
0.7
0.1
5.0
I3
1000
50
1000.0
0.7
0
5.0
N
1000
50
1001.0
0.7
-1
5.0
I1
2000
50
2001
1.40
-1
10
I2
2000
50
2000
1.40
0
10
I3
2000
50
2000
1.40
0
10
N
2000
50
2001
1.40
-1
10
Table 40: MI 2883 current calibration performed with CS 2890 (Uncertainty based
on Calibration certificate)
Содержание CS 2890
Страница 1: ...Power Simulator Calibrator CS 2890 Instruction manual Version 1 4 2 Code No 20 752 950...
Страница 88: ...CS 2890 Power Simulator Calibrator Annex 88...
Страница 89: ...CS 2890 Power Simulator Calibrator Annex 89...
Страница 90: ...CS 2890 Power Simulator Calibrator Annex 90...
Страница 91: ...CS 2890 Power Simulator Calibrator Annex 91...
Страница 92: ...CS 2890 Power Simulator Calibrator Annex 92...
Страница 94: ...CS 2890 Power Simulator Calibrator Annex 94...
Страница 95: ...CS 2890 Power Simulator Calibrator Annex 95...
Страница 96: ...CS 2890 Power Simulator Calibrator Annex 96...