CS 2890 Power Simulator / Calibrator
General setup
17
Selects Colour scheme.
ENTER
Returns to the General setup menu.
3.2 Calibrator
Calibration function is used for
calibrating and adjusting
of Metrel Power Quality
Analysers, like: MI 2883, MI 2885, MI 2892 and MI 2893. Calibrator assures stable
voltage reference for periodic control of the instruments.
Figure 8: Calibrator screen
Table 4: Description of Calibrator options
Instrument name
:
MI 2883, MI 2885, MI 2892 and MI 2893
Fundamental voltage
Select calibration test point for
voltage output (5 ÷ 500 V)
Note: In the calibration mode, voltage is generated via
terminals L1
– L2
Fundamental current
Select voltage calibration point for current output terminals
(0,050 ÷ 2.000 V)
Frequency
Calibration frequency 50.00/60.00 Hz
Table 5: Keys in Calibration menu
F1
Activates adjustment and fine output voltages tuning. See next
Select parameter value
Scrolls cursor between options
Returns to the GENERAL SETUP (one press); MAIN menu
(second press)
Note:
After leaving CALIBRATION menu (test terminals L1
– L2) bi-phase voltage
automatically appears on DUT.
Содержание CS 2890
Страница 1: ...Power Simulator Calibrator CS 2890 Instruction manual Version 1 4 2 Code No 20 752 950...
Страница 88: ...CS 2890 Power Simulator Calibrator Annex 88...
Страница 89: ...CS 2890 Power Simulator Calibrator Annex 89...
Страница 90: ...CS 2890 Power Simulator Calibrator Annex 90...
Страница 91: ...CS 2890 Power Simulator Calibrator Annex 91...
Страница 92: ...CS 2890 Power Simulator Calibrator Annex 92...
Страница 94: ...CS 2890 Power Simulator Calibrator Annex 94...
Страница 95: ...CS 2890 Power Simulator Calibrator Annex 95...
Страница 96: ...CS 2890 Power Simulator Calibrator Annex 96...