CS 2890 Power Simulator / Calibrator
Operating the instrument
58
5.2 Simulation campaign
In following section recommended signal simulation is described. Metrel PQA MI 2892
is used as tested instrument. Refer to Power Master MI 2892 Instruction manual for
handling measuring site. We recommend to strictly follow the guidelines in order to
avoid common problems, measurement and simulation mistakes. Figure below shortly
summarizes recommended simulation practice. Each step is then shortly described in
details.
Figure 62: Recommended simulation practice
5.2.1 Step 1: Connection setup
Preparation of Power Simulator/Calibrator CS 2890 and Power Master MI 2892 includes
the following steps:
Visually check both instruments and accessories.
Make sure, that Power Simulator/Calibrator CS 2890 is turned off.
Содержание CS 2890
Страница 1: ...Power Simulator Calibrator CS 2890 Instruction manual Version 1 4 2 Code No 20 752 950...
Страница 88: ...CS 2890 Power Simulator Calibrator Annex 88...
Страница 89: ...CS 2890 Power Simulator Calibrator Annex 89...
Страница 90: ...CS 2890 Power Simulator Calibrator Annex 90...
Страница 91: ...CS 2890 Power Simulator Calibrator Annex 91...
Страница 92: ...CS 2890 Power Simulator Calibrator Annex 92...
Страница 94: ...CS 2890 Power Simulator Calibrator Annex 94...
Страница 95: ...CS 2890 Power Simulator Calibrator Annex 95...
Страница 96: ...CS 2890 Power Simulator Calibrator Annex 96...