CS 2890 Power Simulator / Calibrator
General setup
21
Note
: Adjust all calibration points from the calibration certificate, which are needed for
the calibration.
2. Adjustment of CS 2890 voltage by using
Substandard Volt-meter
:
Connect substandard Volt-meter to CS 2890:
Figure 13: Connection for the voltage adjustment procedure using substandard Volt-
meter
Connect L1 and L2 outputs from the CS 2890 to substandard volt-meter inputs.
Press ADJUST (1.)
Press
to Enter the Adjustment menu
Press Enter and
enter the corrected value of 5,0045V (value from the substandard volt-meter) and
confirm the value by pressing Enter key