CS 2890 Power Simulator / Calibrator
Operating the instrument
67
6.1.6 Calibration of current inputs for Class A PQA (MI 2892 example)
Reference
value from the CS 2890
UUC
indicated value on the calibration item (MI 2892)
Uncertainty
expanded measurement uncertainty (k=2)
Error
UUC
– Reference
Limit
limit of error (
±), for MI 2892 ±0.25%
UUC set-up: Voltage ratio 1:1, Connection: 4W, Synchronisation: I1, System frequency:
50 Hz, Primary current: 5A, Secondary current: 5A
Reference Frequency
UUC
Uncertainty
Error
Limit
Range=100A
Input
mV
Hz
A
A
A
A
I1
50
50
49.98
0.02
0.02
0.13
I2
50
50
50.01
0.02
-0.01
0.13
I3
50
50
50.02
0.02
-0.02
0.13
N
50
50
50.07
0.02
-0.07
0.13
I1
100
50
99.96
0.03
0.04
0.25
I2
100
50
99.95
0.03
0.05
0.25
I3
100
50
100.0
0.03
0
0.25
N
100
50
100.1
0.03
-0.1
0.25
I1
200
50
199.9
0.06
0.1
0.5
I2
200
50
199.9
0.06
0.1
0.5
I3
200
50
200.0
0.06
0
0.5
N
200
50
200.1
0.06
-0.1
0.5
Range=1000A
I1
50
50
49.99
0.03
0.01
0.13
I2
50
50
50.01
0.03
-0.01
0.13
I3
50
50
50.02
0.03
-0.02
0.13
N
50
50
50.06
0.03
-0.06
0.13
I1
1000
50
1000.0
0.14
0
2.5
I2
1000
50
999.9
0.14
0.1
2.5
I3
1000
50
1000.0
0.14
0
2.5
N
1000
50
1001.0
0.14
-1
2.5
I1
2000
50
2001
1
-1
5
I2
2000
50
2000
1
0
5
I3
2000
50
2000
1
0
5
N
2000
50
2001
1
-1
5
Table 41: MI 2892 current calibration performed with CS 2890 (Uncertainty based
on substandard Volt-meter Keysight 34461)
Содержание CS 2890
Страница 1: ...Power Simulator Calibrator CS 2890 Instruction manual Version 1 4 2 Code No 20 752 950...
Страница 88: ...CS 2890 Power Simulator Calibrator Annex 88...
Страница 89: ...CS 2890 Power Simulator Calibrator Annex 89...
Страница 90: ...CS 2890 Power Simulator Calibrator Annex 90...
Страница 91: ...CS 2890 Power Simulator Calibrator Annex 91...
Страница 92: ...CS 2890 Power Simulator Calibrator Annex 92...
Страница 94: ...CS 2890 Power Simulator Calibrator Annex 94...
Страница 95: ...CS 2890 Power Simulator Calibrator Annex 95...
Страница 96: ...CS 2890 Power Simulator Calibrator Annex 96...