PN T900-014, Rev B
Nitrogen Carrier Gas (for Oxygen and Hydrogen Models Only)
Specification
Description
Inlet Fitting
1/4-inch male face seal fitting*
Inlet pressure range
4.8–6.1 BAR (70–90 PSIG)
Inlet pressure stability
+2 %, UHP regulator required
Flow rate
300 cc/minute
Purity
99.9999 % (<1 PPM total impurity CO between calibration)
* 1/4-inch face seal x 1/16-inch adapter fittings are available from ΑΜΕΤΕΚ
.
Chassis
Specification
Description
Dimensions
18 cm (7") H x 43 cm (16.8") W x 67 cm (26.5") D
Weight
15.9 kg (35 lb)
Power
100–120 VAC, 50/60 Hz
200–240 VAC, 50/60 Hz
Outputs
Specification
Description
Display
LCD graphics, backlit, 100 mm x 150 mm
Printer Port
Concentration, chromatogram, and diagnostic reports
Serial PLC Port
Concentration data and general alarm via RS-232
4–20 mA (optional)
Concentration signals
ta7000F PLC Output Format
The concentration data, status, and other detail information is sent from Port 2
of the ta7000F automatically at the end of every run. The format used is called
Comma Separated Variable (CSV), a common MS Excel format.
The fields within each record are defined below. Fields 1–6 are the record iden-
tification fields and Fields 7–8 are the impurity fields.
The next impurities will be at Fields 9–10, 11–12, 13–14, 15–16, and 17–18. This
string is terminated by a carriage return (CR), and line feed (LF) characters.
The parameters are 9,600 Baud, 8 Data Bits, 1 Stop Bit, and No Parity.
5-4 | ta7000F Gas Purity Monitor