PN T900-014, Rev B
The Warm-Up Cycle typically takes 20–30 minutes. After the Warm-Up Cycle
has completed successfully (the “
Successful Warm-up
” message is displayed)
the
Run
command will appear (Figure 2-7.2). Press the
Run
key to view the
Main Menu
screen (see Figure 2-7.3).
Starting the ta7000F in the Warm-Up Cycle
The Initial start-up screen (see Figure 2-6) will display two menu items:
Warmup
and
Page
. Press
Page
to view the first of three (3) system setup
pages. Refer to “Working From the Page Screens,” later in this chapter. Press the
Warmup
key to initiate the instrument’s
Warm-Up Cycle
. This cycle will place
the instrument in a ready state for analysis.
After pressing the
Warmup
key the ta7000F will begin heating critical zones
and will perform a series of system tests. The results from these tests will ap-
pear on the
Warm-Up
screen. The
Warmup
menu item (bottom of the screen)
will change to
Cool
(Figure 2-7.1). If you wish to stop the
Warm-up Cycle
,
press the
Cool
key and the instrument begin to cool down, and will return to a
completely neutral state. If any errors occur during the Warm-Up Cycle, refer to
“Error Messages” in Chapter 4.
Figure 2-7.1.
Warm-Up Mode screen.
Warmup Mode
12:35:25
Page
Module Electronics: Pass
Detector System:
Detector Temperatures:
Carrier Pressure:
Cool
WARM-UP
Figure 2-7.2.
Warm-Up “Pass” screen.
WARM-UP
Idle Mode
12:34:25
Module Electronics:
Detector System:
Detector Temperatures:
Carrier Pressure:
Pass
Pass
Pass
Pass
***** Successful Warm-up *****
Run Cool
Page
2-18 | ta7000F Gas Purity Monitor