PN T900-014, Rev B
Working From the Calibration Screen
From the
Temperatures & Status
screen press the
Page
key to view the
Cali-
bration
screen (Figure 3-6), which displays information about the most recent
calibration run. From this screen you can also configure Auto-Calibration param-
eters.
Name
This is the gas species to be measured.
Conc.
This is the cylinder concentration (or blend) of the Calibration gas con-
nected to the instrument.
Area
This is the area of the last analysis run for each gas type.
Rfactor
This is the
Response Factor
created by the most recent (or previous) calibra-
tion cycle.
Rfactor
is used to generate the area.
RT
This is the Retention Time, which is the time point of the peak from the
start of the most recent analysis run.
Conc. Factor
This is a value that is preset at the factory. This number represents the
internal gas blender ratio.
Figure 3-6.
Calibration screen.
Calibration
Idle Mode
12:34:25
Edit Page
Name Conc. Area Rfactor RT
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
CH
4
CO
2
NMHC
4800
42842
5300
38236
10000
40571
*
Conc. Factor 200
CalTime 00:00
Cal Purge 1 Min
Interval None
CalTol 10%
3-8 | ta7000F Gas Purity Monitor