Revision 21_Sep_2015
50
User Guide Version 2.60 I-button WHO 5th
APPENDIX 18: Product Performance Data
Abbreviations:
TSC:
Sperm Concentration (Count)
MSC: Motile
Sperm
Concentration
PMSC:
Progressive Motile Sperm Concentration
Morph Norm Forms: Morphologically Normal Forms
OD: Optical
Density
MV: Millivolt
Performance Data Summary
The performance the SQA-V is summarized in the text, tables and graphs below. All values concerning sperm
concentration measurements are expressed as x10
6
sperm cells per milliliter (M/ml). Motility and Morphology values
are expressed as a percent (%). All testing was performed using human patient and donor semen samples.
Calibration:
Each SQA-V is biologically calibrated against two reference systems at Medical Electronic System's laboratory.
Dynamic Range:
Precision and Accuracy Established Against a Known Target
(Latex beads)
Background:
The precision and accuracy of the SQA-V was
compared to a known target value using latex beads (Accu-
beads®).
Latex beads are used as a quality control product to validate
the accuracy of sperm counting methods for two known levels
of concentration. In accordance with CLIA regulations such a
control is used to demonstrate operator proficiency using the
microscope and for validation of automated sperm counting
methods. The latex beads were run in the SQA-V in the same
manner semen samples are run on the system.
Limitations of method:
Latex beads cannot:
Measure sperm motility or morphology
Correct for inaccurate chamber depths or technician
errors
Method comparison:
A total of 320 latex bead samples were tested on ten SQA-V
systems (32 samples/SQA-V). SQA-V concentration readings
were compared to established target /- acceptable
range.
Latex beads established target /- ranges
(Hemacytometer):
Vial #1: 47 +/- 7.0 M/ml
Vial #2: 24 +/- 3.4 M/ml
Precision
SQA-V Accu-beads®
CV,
%
Intra-
device
Variability
High 47± 7.0 M/ml
≤
0.01
Low 24 ± 3.4 M/ml
≤
0.01
Inter-
device
Variability
High 47± 7.0 M/ml
≤
2.00
Low 24 ± 3.4 M/ml
≤
2.50
Accuracy: High Level Control
Accuracy: Low Level Control
SQA-V QUALITY CONTROL TEST
(High level of Accu-beads, target range: 40-54 M/ml)
56
52
48
44
40
36
0
1
2
3
4
5
6
7
8
9
10
Number of SQA-V tested
SQA-V QUALITY CONTROL TEST
(Low level of Accu-beads, target range : 20.6-27.4 M/ml)
30
28
26
24
22
20
18
0
1
2
3
4
5
6
7
8
9
10
Number of SQA-V tested
Ac
c
u
be
a
d
s
co
n
cen
tr
a
ti
o
n
in
mi
lli
o
n
s
pe
r
ml
A
ccu
b
ead
s
conce
n
tr
at
io
n
in
m
ill
io
n
s
pe
r
ml
Sample
Type
Test
Mode
Sperm
Conc.
M/ml
Motility
%
Morph
%
MSC
M/ml
PMSC
M/ml
#Sperm
Cells/field
Fresh Normal 2-400 0-100 0-100
.2-400 0-400
-
Washed Normal
2-200+ 0-100 0-100 .2-200+ 0-200+
-
Frozen Normal
-
-
- .2-200+
0-200+
-
Post-Vasectomy
-
-
- 0-2 - 0-30