User Guide Version 2.60 I-button WHO 5th
Revision 21_Sep_2015
16
If the sample was > 0.5 ml. the screens below will be seen.
The system will first perform an auto-calibration and check for electronic noise.
Do not insert a testing capillary, slide, nor use ANY of the SQA-V keypad or
visualization functions at this time.
PLEA
The S
testing
is plac
testing
The SQA-V algorithm compensates for the sample dilution as long as the sample
has been diluted accurately (If the total sample volume is 0.4 ml then 0.4 ml of a
clear media such as Earle’s buffer must be added).
Recommendation: If the LOW VOLUME sample is viscous, first treat with the
QwikCheck-Liquefaction kit and then dilute the sample for greater accuracy.
Normal
Volume
Samples
SE NOTE:
QA-V will begin
when a capillary
ed into the
chamber.
Testing
FRESH
NORMAL VOLUME SPECIMEN
1.
MIX SAMPLE THOROUGHLY
2.
FILL, CLEAN AND WIPE CAPILLARY
3.
WAIT FOR AUTOCALIBRATION
AUTOCALIBRATION – DO NOT TOUCH UNIT
If the sample was
≥
0.5 ml the screen above will provide instructions for preparing
a testing capillary.
Fill the SQA-V testing capillary according to the instructions in the Appendix
section of this user guide: “Filling the SQA-V Capillary with a Normal Volume
Sample”.
FRESH
NORMAL VOLUME SPECIMEN
1.
MIX SAMPLE THOROUGHLY
2.
FILL, CLEAN AND WIPE CAPILLARY
3.
WAIT FOR AUTOCALIBRATION
INSERT CAPILLARY INTO CHAMBER
The screen above will be displayed when it is time to insert the filled testing
capillary in the measurement compartment, testing will begin automatically.
A sample is tested in approximately 75 seconds. If the sample is low quality, the system
will perform an additional 2 minute test:
TESTING
DO NOT MOVE CAPILLARY OR
OPERATE DEVICE DURING TESTING
TESTING
LOW QUALITY SAMPLE
TESTING WILL TAKE 2 MORE MINUTES