SECTION 9: Operating the Visualization System (Video Display)
The SQA-V Visualization System with video display (upper screen) is used to view and
count sperm cells. The visualization system is a critical "link" to V-Sperm GOLD where
enhanced, real time video can be displayed on a PC monitor. The visualization system:
Accommodates both an SQA-V testing capillary to "scan" through a depth of
300 microns or a standard slide to view samples (20 micron depth).
Operates via control knobs to set focus, brightness, contrast and color, and via
the keypad zoom, illumination, and monitor on/off functions.
Magnification range: x300 to x500.
Standard Slide Preparation:
Use 10 µl of semen.
Standard slide, 22 mm x 22 mm cover-slip (to insure 20 micron depth).
Load the prepared, standard slide into the SQA-V slide adaptor.
Testing Capillary Preparation:
Fill the SQA-V testing capillary for either a normal or low volume specimen
(see Appendix).
Visualization Process:
The video display will automatically illuminate when the SQA-V is turned on.
Use monitor
ON/OFF
key on the keypad to independently operate the video
display.
Wait for the self-test to complete (system is disabled at this time).
To ensure that the visualization system is working properly prior to use:
Press the
HIGH ILLUMINATION
key multiple times to ensure a
maximum level setting.
To view cells:
Press
ZOOM IN
to maximum magnification (x500).
To count cells:
Press
ZOOM OUT
to minimum magnification (x300).
Insert semen sample (either capillary or slide) into the visualization chamber.
Adjust
CONTRAST, COLOR, BRIGHTNESS, FOCUS
and object
ILLUMINATION
controls for optimal image quality.
Use
ZOOM OUT
(x300) /
ZOOM IN
(x500) to regulate magnification.
Counting Cells Using the Visualization Screen:
1.
Follow the WHO Manual instructions for semen sample collection and
preparation. Thoroughly mix the sample before step #2.
2.
Pipette 10uL of the semen sample onto a standard slide and cover with a
22x22 mm coverslip. Prepare a new slide if air bubbles or liquid spillage
occurs.
3.
Load the slide into the slide adaptor and then insert the slide adaptor into the
SQA-V visualization chamber. (Refer to the SQA-V User Guide APPENDIX 3:
Using Standard Slides in the Visualization System for details).
4.
Press the ZOOM-OUT button on the SQA-V keypad all the way to set the
magnification to x300.