Value Description
1
Waiting for controller address
2
Addressed, Buffering chars
3
Message received, not processed yet
States, TX
Transmitter Driver State
Value Description
0 Transmitter
passive
1
Transmitter active
2
The last byte of the telegram is being transmitted
Clean Off Screen
Displays the Start-up Screen for one minute. Since there are no buttons in this screen,
you can safely clean the touch screen.
Calibrate Diag
This screen displays the calibration history for the four last calibration occurrences. You
can see current and previous Zero Weight (
Z ld
) and Scale Factor (
Scl F
); analog with the
date and time they were updated by calibration procedures. Zero Load updates when
you run a Zero Procedure. Scale Factor updates when you run a Weight, Material or
ECal procedure. See Calibrating Your Controller on page 46.
NOTE:
If you change the Zero Weigh or Scale Factor manually in the Calibrate Settings menu
(page 28), there will be no record of the changes in this screen.
Analog Diagnostics
The top section displays the selected function (Cntrl Spd, Batch Tot etc.) and type (0-20
mA or 4-20 mA) for each of the Analog outputs. The
Actual
column in the output section
displays the engineering value, used to generate the output. The
EST-mA
column
displays the estimated current output in mA for the output. See Analog Outputs on page
32 for more information. As long as standard output scaling is used, the translation
between engineering value and output current is straightforward. If you have to use non-
standard output scaling, this is where you can see the effect of your changes to the
analog output scaling. If you are using voltage output, divide
EST-mA
with two to get
Volts.
35.00.EX O&M Manual
59