Set the signal at the source for full-scale value, typically 20 mA. The live counts should
follow the input, and be around 1,000,000 counts. Push the Snap Full button.
The live value will snap to the New column, Hi row.
If the values are acceptable, push the
Accept
button, followed by the
Cancel
button, else
the
Cancel
button only.
35.00.EX O&M Manual
53